Predictive Vulnerability Intelligence.

Product

  • Product
  • Pricing
  • Documentation

Company

  • About
  • Partnerships
  • Blog
  • Support

Legal

  • Terms
  • Privacy
  • Data Licensing

© 2026 FAUCET Technologies LLC. All rights reserved.

CVE-2017-9657

21
FAUCET Score

CVE-2017-9657 describes a vulnerability in the Philips IntelliVue MX40 WLAN monitor (Version B.06.18) where specific 802.11 network conditions can lead to a partial re-association. This can cause the central monitoring station to incorrectly indicate the MX40 is disconnected, while the device itself remains in telemetry mode (audio/screen off), preventing local alarming. Rated 6.5 (Medium) on the CVSS v3 scale, this vulnerability has an adjacent network attack vector with low complexity, and its primary impact is a high availability risk due to potential delays in patient treatment. There is no evidence of active exploitation, public exploit code, or significant community discussion surrounding this CVE.

Impacted Technologies

VendorProductVersion(s)CPE
< b.06.18CPE matchmatch criteria
cpe:2.3:o:philips:intellivue_mx40_firmware:*:*:*:*:*:*:*:*

CVSS Data

CVSS version used by this source: 3.0

6.5MEDIUM

CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

Attack Vector
ADJACENT_NETWORK
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality Impact
NONE
Integrity Impact
NONE
Availability Impact
HIGH
Exploitability Score
2.8
Impact Score
3.6
CvssVersion
3.0

Exploit Intelligence

EPSS Score
0.79%
Probability of exploitation in next 30 days
EPSS Percentile
52.6%
Percentile rank of EPSS score among Peer Group
As of 2026-07-26
Model: v2026.06.15
This CVE's current EPSS score of 0.0079 is in the 82nd percentile among its peer group of 1,802 CVEs.

Social Chatter

The average CVE in this peer group has 0.0 Twitter, 0.0 Reddit, 0.1 Bluesky, 0.0 Mastodon, and 0.1 GitHub mentions.

Media Mentions

No media coverage found for this CVE.

The average CVE in this peer group has 0.1 InfoSec Media, 0.0 Vendor Blog, and 0.0 Security Researcher mentions.

Remediation

Remediation records are not available for this CVE.

References

ics-cert.us-cert.gov / advisories/ICSMA-17-255-01
Third Party AdvisoryUS Government Resource
usa.philips.com / healthcare/about/customer-support/product-security
Vendor Advisory
securityfocus.com / bid/100813
Third Party AdvisoryVDB Entry