CVE-2023-47262 describes a vulnerability in Abbott ID NOW devices prior to v7.1, where physical access to an internal serial port allows for interruption or modification of the startup process and device configurations. This medium-severity vulnerability (CVSS 5.2) requires direct physical access, making it difficult to exploit, and could lead to high integrity and low availability impacts. There is no evidence of active exploitation, public exploit code, or significant community discussion or media coverage surrounding this CVE.
| Vendor | Product | Version(s) | CPE |
|---|---|---|---|
< 7.1CPE matchmatch criteria | cpe:2.3:o:abbott:id_now_firmware:*:*:*:*:*:*:*:* |
CVSS version used by this source: 3.1
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:L
The average CVE in this peer group has 0.0 Twitter, 0.1 Reddit, 0.1 Bluesky, 0.0 Mastodon, and 0.1 GitHub mentions.
No media coverage found for this CVE.
The average CVE in this peer group has 0.2 InfoSec Media, 0.0 Vendor Blog, and 0.0 Security Researcher mentions.
Remediation records are not available for this CVE.