CVE-2020-25187

28
FAUCET Score

CVE-2020-25187 describes a critical heap overflow vulnerability in the Medtronic MyCareLink Smart 25000 Patient Reader software. An authenticated attacker can exploit this by sending a specially crafted debug command, potentially leading to remote code execution and full control of the device. With a CVSS score of 9.8 (CRITICAL), this vulnerability poses a significant risk due to its network-based attack vector, low attack complexity, and high impact on confidentiality, integrity, and availability. While there is no evidence of active exploitation, public exploit code, or KEV listing, the vulnerability has received some community discussion and media coverage, indicating awareness within the cybersecurity community.

Impacted Technologies

VendorProductVersion(s)CPE
All Versions ImpactedCPE matchmatch criteria
cpe:2.3:o:medtronic:mycarelink_smart_model_25000_firmware:*:*:*:*:*:*:*:*

CVSS Data

CVSS version used by this source: 3.1

8.8HIGH

CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H

Attack Vector
ADJACENT_NETWORK
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
REQUIRED
Scope
CHANGED
Confidentiality Impact
HIGH
Integrity Impact
HIGH
Availability Impact
HIGH
Exploitability Score
2.1
Impact Score
6.0
CvssVersion
3.1

Exploit Intelligence

EPSS Score
3.77%
Probability of exploitation in next 30 days
EPSS Percentile
88.8%
Percentile rank of EPSS score among Peer Group
As of 2026-07-27
Model: v2026.06.15
This CVE's current EPSS score of 0.0377 is in the 81st percentile among its peer group of 36,897 CVEs.

Social Chatter

The average CVE in this peer group has 0.0 Twitter, 0.1 Reddit, 0.3 Bluesky, 0.3 Mastodon, and 2.4 GitHub mentions.

Media Mentions

The average CVE in this peer group has 0.3 InfoSec Media, 0.0 Vendor Blog, and 0.0 Security Researcher mentions.

Remediation

Remediation records are not available for this CVE.

References

us-cert.cisa.gov / ics/advisories/icsma-20-345-01
Third Party AdvisoryUS Government Resource
global.medtronic.com / xg-en/product-security/security-bulletins/mycarelink-smart-security-vulnerability-patch.html
cisa.gov / news-events/ics-medical-advisories/icsma-20-345-01