CVE-2014-5433 describes a critical vulnerability in the Baxter SIGMA Spectrum Infusion System (version 6.05 with WBM version 16) where an unauthenticated remote attacker can execute commands to view wireless account credentials stored in cleartext. This allows for potential unauthorized access to the host network. The vulnerability has a CVSS score of 9.8 (CRITICAL) due to its network attack vector, low complexity, and high impact on confidentiality, integrity, and availability. Despite its severity, there is no known active exploitation, public exploit code (Metasploit, Nuclei, ExploitDB), or significant community discussion or media coverage surrounding this CVE. Baxter has released Version 8 of the infusion system to address this and other issues.
| Vendor | Product | Version(s) | CPE |
|---|---|---|---|
6.05CPE matchmatch criteria | cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:* |
CVSS version used by this source: 3.0
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
The average CVE in this peer group has 0.0 Twitter, 0.1 Reddit, 0.3 Bluesky, 0.3 Mastodon, and 2.4 GitHub mentions.
No media coverage found for this CVE.
The average CVE in this peer group has 0.3 InfoSec Media, 0.0 Vendor Blog, and 0.0 Security Researcher mentions.
Remediation records are not available for this CVE.